ANNEX IV

CONTAINMENT AND OTHER PROTECTIVE MEASURES

General principles

1.

These tables present the normal minimum requirements and measures necessary for each level of containment. Containment is also achieved through the use of good work practices, training, containment equipment and special installation design. For all activities involving GMMs the principles of good microbiological practice and the following principles of good occupational safety and hygiene shall apply:

to keep workplace and environmental exposure to any GMM to the lowest practicable level;

to exercise engineering control measures at source and to supplement these with appropriate personal protective clothing and equipment when necessary;

to test adequately and maintain control measures and equipment;

to test, when necessary, for the presence of viable process organisms outside the primary physical containment;

to provide appropriate training of personnel;

to establish biological safety committees or subcommittees, if required;

to formulate and implement local codes of practice for the safety of personnel, as required;

where appropriate, to display biohazard signs;

to provide washing and decontamination facilities for personnel;

to keep adequate records;

to prohibit eating, drinking, smoking, applying cosmetics or the storing of food for human consumption in the work area;

to prohibit mouth pipetting;

to provide written standard operating procedures where appropriate to ensure safety;

to have effective disinfectants and specified disinfection procedures available in case of spillage of GMMs;

to provide safe storage for contaminated laboratory equipment and materials, when appropriate.

2.

The titles of the tables are indicative: Table I A presents minimum requirements for laboratory activities. Table I B presents additions to and modifications of Table I A for glasshouse/growth-room activities involving GMMs. Table I C presents additions to and modifications of Table I A for activities with animals involving GMMs. Table II presents minimum requirements for activities other than laboratory activities. In some particular cases, it might be necessary to apply a combination of measures, from Table I A and Table II, of the same level. In some cases users may, with the agreement of the competent authority, not apply a specification under a particular containment level or combine specifications from two different levels. In these tables ‘optional’ means that the user may apply these measures on a case-by-case basis, subject to the assessment referred to in Article 4(2).

3.

In implementing this Annex, Member States may in addition incorporate in the following tables the general principles set out in points 1 and 2, with a view to clarifying the requirements.

Table I A

Containment and other protective measures for laboratory activities

Specifications

Containment levels

1

2

3

4

1

Laboratory suite: isolation (1)

Not required

Not required

Required

Required

2

Laboratory: sealable for fumigation

Not required

Not required

Required

Required

Equipment

3

Surfaces resistant to water, acids, alkalis, solvents, disinfectants and decontamination agents, and easy to clean

Required (bench)

Required (bench)

Required (bench, floor)

Required (bench, floor, ceiling, walls)

4

Entry to lab via airlock (2)

Not required

Not required

Optional

Required

5

Negative pressure relative to the pressure of the immediate environment

Not required

Not required

Required except for (3)

Required

6

Extract and input air from the laboratory should be HEPA (4)-filtered

Not required

Not required

Required (HEPA — extract air except for (3))

Required (HEPA— input and extract air (5))

7

Microbiological safety post

Not required

Optional

Required

Required

8

Autoclave

On site

In the building

En suite (6)

In lab = double-ended

System of work

9

Restricted access

Not required

Required

Required

Required

10

Biohazard sign on the door

Not required

Required

Required

Required

11

Specific measures to control aerosol dissemination

Not required

Required minimise

Required prevent

Required prevent

13

Shower

Not required

Not required

Optional

Required

14

Protective clothing

Suitable protective clothing

Suitable protective clothing

Suitable protective clothing and (optional) footwear

Complete change of clothing and footwear before entry and exit

15

Gloves

Not required

Optional

Required

Required

18

Efficient vector control (e.g. for rodents and insects)

Optional

Required

Required

Required

Waste

19

Inactivation of GMMs in effluent from hand-washing sinks or drains and showers and similar effluents

Not required

Not required

Optional

Required

20

Inactivation of GMMs in contaminated material and waste

Optional

Required

Required

Required

Other measures

21

Laboratory to contain its own equipment

Not required

Not required

Optional

Required

23

An observation window or alternative is to be present so that occupants can be seen

Optional

Optional

Optional

Required

Table I B

Containment and other protective measures for glasshouses and growth-rooms

The terms ‘glasshouse’ and ‘growth-room’ refer to a structure with walls, a roof and a floor designed and used principally for growing plants in a controlled and protected environment.

All provisions of Table I A shall apply with the following additions/modifications:

Specifications

Containment levels

1

2

3

4

Building

1

Glasshouse: permanent structure (7)

Not required

Required

Required

Required

Equipment

3

Entry via a separate room with two interlocking doors

Not required

Optional

Optional

Required

4

Control of contaminated run-off water

Optional

Minimise (8) run-off

Prevent run-off

Prevent run-off

System of work

6

Measures to control undesired species such as insects, rodents, arthropods

Required

Required

Required

Required

7

Procedures for transfer of living material between the glasshouse/growth-room, protective structure and laboratory shall control dissemination of GMMs

Minimise dissemination

Minimise dissemination

Prevent dissemination

Prevent dissemination

Table I C

Containment and other protective measures for activities in animal units

All provisions of Table I A shall apply with the following additions/modifications:

Specifications

Containment levels

1

2

3

4

Facilities

1

Isolation of animal unit (9)

Optional

Required

Required

Required

2

Animal facilities (10) separated by lockable doors

Optional

Required

Required

Required

3

Animal facilities designed to facilitate decontamination (waterproof and easily washable material (cages, etc.))

Optional

Optional

Required

Required

4

Floor and/or walls easily washable

Optional

Required (floor)

Required (floor and walls)

Required (floor and walls)

5

Animals kept in appropriate containment facilities such as cages, pens or tanks

Optional

Optional

Optional

Optional

6

Filters on isolators or isolated room (11)

Not required

Optional

Required

Required

Table II

Containment and other protective measures for other activities

Specifications

Containment levels

1

2

3

4

General

1

Viable micro-organisms should be contained in a system which separates the process from the environment (closed system)

Optional

Required

Required

Required

2

Control of exhaust gases from the closed system

Not required

Required, minimise dissemination

Required, prevent dissemination

Required, prevent dissemination

3

Control of aerosols during sample collection, addition of material to a closed system or transfer of material to another closed system

Optional

Required, minimise dissemination

Required, prevent dissemination

Required, prevent dissemination

4

Inactivation of bulk culture fluids before removal from the closed system

Optional

Required, by validated means

Required, by validated means

Required, by validated means

5

Seals should be designed so as to minimise or prevent release

No specific requirement

Minimise dissemination

Prevent dissemination

Prevent dissemination

6

The controlled area should be designed to contain spillage of the entire contents of the closed system

Optional

Optional

Required

Required

7

The controlled area should be sealable to permit fumigation

Not required

Optional

Optional

Required

Equipment

8

Entry via airlock

Not required

Not required

Optional

Required

9

Surfaces resistant to water, acids, alkalis, solvents, disinfectants and decontamination agents, and easy to clean

Required (bench if any)

Required (bench if any)

Required (bench if any, floor)

Required (bench, floor, ceiling, walls)

10

Specific measures to adequately ventilate the controlled area in order to minimise air contamination

Optional

Optional

Optional

Required

11

The controlled area should be maintained at an air pressure negative to the immediate surroundings

Not required

Not required

Optional

Required

12

Extract and input air from the controlled area should be HEPA filtered

Not required

Not required

Required (extract air, optional for input air)

Required (input and extract air)

System of work

13

Closed systems should be located within a controlled area

Not required

Optional

Required

Required

14

Access should be restricted to nominated personnel only

Not required

Required

Required

Required

15

Biohazard signs should be posted

Not required

Required

Required

Required

17

Personnel should shower before leaving the controlled area

Not required

Not required

Optional

Required

18

Personnel should wear protective clothing

Required (work clothing)

Required (work clothing)

Required

Complete change before exit and entry

Waste

22

Inactivation of GMMs in effluent from hand-washing sinks and showers or similar effluents

Not required

Not required

Optional

Required

23

Inactivation of GMMs in contaminated material and waste, including those in process effluent before final discharge

Optional

Required, by validated means

Required, by validated means

Required, by validated means

(1)

Isolation= the laboratory is separated from other areas in the same building or is in a separate building.

(2)

Airlock= entry must be through an airlock which is a chamber isolated from the laboratory. The clean side of the airlock must be separated from the restricted side by changing or showering facilities and preferably by interlocking doors.

(3) Activities where transmission does not occur via airborne route.

(4)

HEPA= High efficiency particulate air.

(5) Where viruses which are not retained by HEPA filters are used, extra requirements will be necessary for extract air.

(6) With validated procedures, allowing the safe transfer of material into an autoclave outside the lab, and providing an equivalent level of protection.

(7) The glasshouse shall consist of a permanent structure with a continuous waterproof covering, located on a site graded to prevent entry of surface-water run-off, and with self-closing lockable doors.

(8) Where transmission can occur through the ground.

(9)

Animal unit: a building or separate area within a building containing facilities and other areas such as changing rooms, showers, autoclaves, food storage areas, etc.

(10)

Animal facility: a facility normally used to house stock, breeding or experimental animals or one which is used for the performance of minor surgical procedures.

(11)

Isolators: transparent boxes where small animals are contained within or outside a cage; for large animals, isolated rooms may be more appropriate.